Probe Care
The OxyGuard profile will give you many years of trouble-free service if you treat it with a little care. The membrane should be kept clean, and the wooden insert must not be allowed to dry out. Check that the sponge in the protection cap is wet every time you use the probe, and if you store it for any length of time check that the sponge is wet at regular intervals. Calibrate the probe as needed - once a day should be more than enough. Renovate the probe if you cannot calibrate to the correct value.
Probe Calibration
The Probe should be calibrated in water-saturated air, or in air-saturated water. We recommend the following procedure:
1) Unscrew the protection cap and remove the sponge.
2) Wipe the membrane - it should be clean and dry.
3) Put a few drops of water into the protection cap and lightly screw it in place.
4) Adjust the electronics connected to give the calibration value corresponding to 100% sat.
Probe Renovation
1) Clean the outside of the probe. Unscrew the membrane cap, discard the used membrane and O-ring. Clean and dry the cap.
2) Soak the probe end in electrolyte - how long depends on how long ago it was last soaked.
3) Inspect the cathode face. If the wooden insert has lifted above the cathode or edge wet the grade 300 wet-or-dry emery paper delivered with the probe with electrolyte and use it to smooth it down until no edges can be felt, using the electrolyte as lubricant.
4) Put a new O-ring, well greased with silicone grease, in place in the bottom of the cap and place a new membrane above it. It is very important that the membrane is placed concentrically in the bottom of the cap above the O-ring. Take care not to get grease on the membrane.
5) Dip the cathode face in electrolyte.
6) Hold the probe with the cathode face up and gently screw the cap with O-ring and membrane onto it. Tighten the cap firmly. The membrane should not wrinkle, if it does discard it and start again with a new membrane.
7) Renovation is complete. Calibrate the probe. Remember to store with the protector fitted. The sponge in the protector must be kept moist at all times.