1 |
Fluids |
Types |
Any Homogeneous and Sonically Conductive Liquid |
| Temperature range |
-4 to 250 °F |
| Turbidity |
10,000 mg/l or less recommended for optimum performance |
|
2 |
Pipes |
Types |
Carbon steel, stainless steel, cast iron, plastic, copper, ductile, etc. |
| Diameter |
*Standard transducers = 2 to 40 inches Small transducers for = 1/2 to 8 inches Large transducers for = 12 to 196 inches (For larger pipes please consult factory) |
| Lining |
Tar, Epoxy Resins, Metal, etc. |
| Minimum Straight pipe |
Upstream side = 10 X Pipe diameter Downstream side = 5 X Pipe diameter |
|
3 |
Measuring range |
Velocity |
-65 to +65 feet/second |
|
4 |
Accuracy |
|
Pipe Diameter |
Flow Rate > 3 feet/sec |
Flow Rate < 3 feet/sec |
|
1/2 to 8 inch 2 to 40 inch 12 to 196 inch |
±2.0% ±1.5% ±1.0% |
±0.787 inches/sec ±0.591 inches/sec ±0.394 inches/sec | |
|
5 |
Measuring method |
Transit time |
|
6 |
Output signal |
Analog output Output items Single output format Dual output format |
Instantaneous flow rate 4 ~ 20 mA DC (600 ohm or less, non-isolated) (selectable) 1 ~ 4 ~ 20 mA DC (600 ohm or less, non-isolated) (selectable) |
Digital output output items output format |
Measuring time, instantaneous/ rate, integrated contra-flow rate, error codes(selectable) Via RS232C interface |
Printer output output items |
Measuring time, instantaneous/ rate, integrated contra-flow rate, error codes(selectable)
|
|
7 |
Display |
Content |
Instantaneous flow rate/instantaneous velocity 4 digits, integrated flow rate 6 digits, flow/contra-flow (trend graph), item setting and data set, self-diagnosis results. |
| Format |
128 X 128 Dot matrix |
|
8 |
Settings |
Method |
Interactive key input |
| Items |
Pipe data settings, analog output, digital output, etc. |
|
9 |
Wall Thickness measurement |
Range Setting range |
0.06 to inches (steel) 1640 to 32,805 feet/sec |
|
10 |
Speed of Sound Range |
Range |
328 to 9842 feet/sec |
|
11 |
Memory back up |
Built-in lithium battery, integrated flow rate data, measurement setting data, logging data and measured data for 5 years |
|
12 |
Logging function |
Stores approx. 55,000 items (measuring time, instantaneous rate, integrated rate, integrated contra rate, error codes |
|
13 |
Power Supply |
Power source 11~ 30 V DC |
| Battery 7.2 V DC (7.5 hours) |
| AC adapter |
output voltage input voltage |
12V DC @ 11 VA * Battery recharge time ~ 4 hours. 1. 100/120 VAC 50/60 Hz (selectable) | |
|
14 |
Power consumption |
Approx. 3 Watts (on battery) Approx. 10 Watts (12 VDC include battery recharging) |
|
15 |
Temperature range |
-15 ~ +125 °F |
|
16 |
Humidity |
90% RH or less (no condensation) |
|
17 |
Weight |
Approx. 2.8 pounds (with battery) |